ENDURANT STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2012-01954
- Event Type
- Injury
- Date Received
- October 16, 2012
- Date of Event
- September 18, 2012
- Report Date
- September 18, 2012
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: INHERENT RISK OF PROCEDURE (OCCLUSION); INCORRECT TECHNIQUE/PROCEDURE (NOT PROPERLY VISUALIZING THE HYOGASTRIC ARTERY). EVALUATION, CONCLUSION: OPERATIONAL CONTEXT CONTRIBUTED TO EVENT (NOT PROPERLY VISUALIZING THE HYOGASTRIC ARTERY).
AN ENDURANT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 4.9 CM IN DIAMETER ABDOMINAL AORTIC ANEURYSM APPROXIMATELY ONE MONTH AGO. VESSEL MORPHOLOGY WAS REPORTED AS THE PROXIMAL AORTA WAS 18.2-22.3 MM IN DIAMETER AND 32 MM IN LENGTH. THE RIGHT ILIAC ARTERY WAS 9.7-9.4-10.0 MM IN DIAMETER AND THE LEFT ILIAC ARTERY WAS 8.5-10.9-10.4 MM IN DIAMETER. THE RIGHT FEMORAL ARTERY WAS 7.8 MM IN DIAMETER AND THE LEFT FEMORAL ARTERY WAS 8.1 MM IN DIAMETER. BOTH ILIAC ARTERIES HAD CALCIFICATION. THE DISTANCE FROM THE BIFURCATION TO THE HYPOGASTRIC WAS 164.5 MM. AFTER THE BIFURCATED STENT GRAFT WAS DEPLOYED, IT WAS NOTICED THAT THE LEFT INTERNAL ILIAC ARTERY WAS UNINTENTIONALLY COVERED. THE COVERAGE WAS DUE TO INCORRECTLY IDENTIFYING THE LOCATION OF THE LEFT INTERNAL ILIAC ARTERY WITH IMAGING; THE ARTERY WAS VISUALIZED WHILE THE DELIVERY SYSTEM WAS OCCUPYING THE SAME SPACE. THE FINAL ANGIOGRAM SHOWED RAPID COLLATERALIZATION OF THE INTERNAL ILIAC ARTERY AND THEREFORE THE PHYSICIAN EXPECTS THE PATIENT TO BE NON-SYMPTOMATIC. THE PATIENT STATUS POST-PROCEDURE WAS FINE. NO CLINICAL SEQUELAE WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDURANT STENT GRAFT SYSTEM | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V01076204 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00063 YR |