LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2012-04749
- Event Type
- Malfunction
- Date Received
- October 16, 2012
- Date of Event
- October 3, 2012
- Report Date
- October 12, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). ADDITIONAL INFORMATION: WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON?---AFTER SEVERAL FIRINGS. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? ---BLOOD VESSEL. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? ---NO. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? ---NO INFORMATION. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? ---NO INFORMATION. WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? ---NO INFORMATION. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? ---NO INFORMATION. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? ---NO INFORMATION. WHAT WERE THE INDICATIONS FOR SURGERY? WHAT WAS FOUND? ---NO INFORMATION. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? IF SO, WHAT? ---NO INFORMATION. DID SOMEBODY OTHER THAN THE PRIMARY SURGEON FIRE THE INSTRUMENT? IF SO, WHOM? ---NO INFORMATION. THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING THE DEVICE FOR FUNCTIONALITY THE DEVICE WAS CYCLED, FED, AND FORMED SIX CLIPS AS INTENDED AND IT EJECTED ONE CLIP. FINALLY, IT LOCKED OUT AS INTENDED. PLEASE NOTE THAT AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE DROPPING/EJECTED CLIPS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIP EJECTED AFTER SEVERAL FIRINGS. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | H43Z14 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |