EON
Report
- Report Number
- 1627487-2012-01835
- Event Type
- Injury
- Date Received
- October 10, 2012
- Date of Event
- September 19, 2012
- Report Date
- September 19, 2012
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2012-1836. IT WAS REPORTED THE PT IS EXPERIENCING A HEATING SENSATION AT HER IPG SITE WHILE CHARGING. SHE IS REQUESTING HER SCS SYSTEM TO BE REMOVED. F/U INFO IDENTIFIED THE PT'S ISSUE WAS RESOLVED BY CHARGING FOR SHORTER TIME PERIODS. ON (B)(6) 2012 ST. JUDE MEDICAL, NEUROMODULATION DIV., SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EON | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3716 | 55569 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Required Intervention | SCS LEAD: MODEL 3186 (2)| IMPLANT: |