INFINITI VISION SYSTEM OZIL
Report
- Report Number
- 2028159-2012-01579
- Event Type
- Injury
- Date Received
- October 12, 2012
- Date of Event
- September 12, 2012
- Report Date
- September 12, 2012
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NURSE
Narratives
THE FACILITY DID NOT REQUEST SERVICE. HOWEVER, THE CUSTOMER RETURNED A PHACO HANDPIECE FOR EVALUATION. THE HANDPIECE (HP) WAS EXAMINED AND A VISUAL ASSESSMENT OF THE SAMPLE REVEALED A DENTED IRRIGATION LINE AND A SEPARATION OF BOTH STRAIN RELIEFS. THE HP GROUND RESISTANCE, U/S STROKE LENGTH, AND TORSIONAL STROKE LENGTH WERE MEASURED AND FOUND TO BE WITHIN PRODUCT SPECIFICATIONS. THE IRRIGATION AND ASPIRATION PASSAGE FLUIDIC FLOW TEST WAS PERFORMED AND THE FLOW RATES WERE FOUND TO MEET PRODUCT SPECIFICATIONS. THE HP SUCCESSFULLY PRIMED AND TUNED ON A CALIBRATED SYSTEM AND PERFORMED AT 100% PHACO POWER FOR 5 MINUTES WITHOUT ANY OBSERVABLE ISSUES. AN EVALUATION OF THE RETURNED SAMPLE REVEALED SEVERAL COSMETIC NONCONFORMITIES. THERE WERE DENTS ON THE IRRIGATION LINE AND A SEPARATION OF BOTH THE HP AND CONNECTOR STRAIN RELIEFS; HOWEVER, THESE ISSUES ARE NOT RELATED TO THE REPORTED EVENT. THE HP PASSED ALL FUNCTIONAL TESTING AND WAS VERIFIED TO MEET PRODUCT SPECIFICATIONS. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS PHACO HANDPIECE OR SYSTEM. CORNEAL BURN IS AN ISSUE THAT IS OCCASIONALLY REPORTED WITH CATARACT SURGERY. ACCORDING TO THE PENNSYLVANIA PATIENT SAFETY ADVISORY ABSTRACT: PREVENTING CORNEAL BURNS DURING PHACOEMULSIFICATION, MARCH 2010, VOL. 7, NO. 1:23-25, MOST CORNEAL BURNS CAN BE TRACED TO ISSUES RELATED TO SURGICAL TECHNIQUE AND NOT TO MALFUNCTIONING EQUIPMENT. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).
A NURSE REPORTED THAT DURING A PROCEDURE, OCCLUSIONS KEPT OCCURRING AND A PATIENT EXPERIENCED A THERMAL BURN. THE CASE WAS COMPLETED USING AN ALTERNATE HANDPIECE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | INFINITI OZIL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | OZIL TORSIONAL PHACO HANDPIECE |