FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2792098 · Received October 11, 2012

Report

Report Number
1627487-2012-01847
Event Type
Injury
Date Received
October 11, 2012
Date of Event
September 20, 2012
Report Date
September 20, 2012
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVALUATION: RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2012-1846. IT WAS REPORTED, THE PT IS EXPERIENCING SEVERE HEATING AT HIS IPG SITE WHILE CHARGING WHEN LAYING DOWN. THE PT STATED CHARGING STANDING UP HAS HELPED THE PROBLEM. THE PT WAS ADVISED TO CHARGE MORE FREQUENTLY FOR SHORTER PERIODS OF TIME. ON 08/01/2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO THE PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CHARGING SYSTEM SCS CHARGER LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3657495

Patients

Seq Age Sex Outcome Treatment
1 35 YR Required Intervention IMPLANTED:| SCS ANCHOR: MODEL 1192| SCS LEAD: MODEL 3186| IMPLANTED: