FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2791875 · Received October 15, 2012

Report

Report Number
3004209178-2012-90106
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
August 22, 2012
Report Date
October 4, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED AN ERROR DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAD AN ERROR ALARM DURING THE MANUAL PRIME PROCESS. ADVISED THE CALLER TO REVERT TO BACK UP PLAN. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 500MG/DL, AND SHE WAS TREATED WITH MANUAL INJECTIONS. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 17 YR