FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2791833 · Received October 15, 2012

Report

Report Number
2032227-2012-07398
Event Type
Injury
Date Received
October 15, 2012
Date of Event
September 27, 2012
Report Date
September 30, 2012
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 595MG/DL, AND SHE WAS TREATED WITH MANUAL INJECTIONS. IT WAS STATED THAT THE CUSTOMER EXPERIENCED NAUSEA AND VOMITING PRIOR HER ADMISSION. TROUBLESHOOTING WAS PERFORMED. THE TIME, DATE, AND SETTINGS WERE CORRECT. ASSISTED THE CALLER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED THE HIGH PRESSURE TEST AND PASSED. INSTRUCTED TO CHANGE THE INFUSION SET AND RESERVOIR. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC MINIMED MMT-522LNAB

Patients

Seq Age Sex Outcome Treatment
1 19 YR Hospitalization