CAPSUREFIX NOVUS MRI
Report
- Report Number
- 2649622-2012-14557
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- August 7, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. HOWEVER, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT THE LEAD SCREW DID NOT COME OUT. IT WAS NOTED THAT AFTER APPROXIMATELY TWENTY TURNS THE SCREW SUDDENLY "POPPED OUT" AND DISLOCATED FROM THE HEART WALL BECAUSE OF THE FORCE RELEASED. IT WAS ALSO REPORTED THAT ON THE X-RAY IT LOOKED LIKE THE SPACERS WERE "STUCK" AND THEY DID NOT MOVE AT ALL DURING TURNING. THE LEAD WAS NOT USED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |