SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2012-15159
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- Z-0474-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR FRACTURED. IT WAS NOTED THAT THE DEFIBRILLATION CONDUCTOR WAS DISTORTED, THE DEFIBRILLATION CONDUCTOR WAS CUT, THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER TUBING OVERLAY WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD) AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL BECAUSE A LEAD INTEGRITY ALERT [LIA] HAD BEEN TRIGGERED. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR [RV] LEAD WAS OVERSENSING, NOISE WAS PRESENT AND THE LEAD WAS FRACTURED. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |