FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2791741 · Received October 15, 2012

Report

Report Number
2649622-2012-14863
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
August 21, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVY
PMA / PMN Number
P920015/S055
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX/LOBE WAS DISTORTED/BENT. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THE TURNS TO EXTEND/RETRACT THE HELIX EXCEEDS SPECIFICATION, AND THE LEAD APPEARED DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT OF THE RIGHT VENTRICULAR LEAD, THE HELIX WAS EXTENDED THEN RETRACTED IN ORDER TO BE REPOSITIONED. THE HELIX WOULD NOT EXTEND AGAIN. THE LEAD WAS REMOVED FROM THE PATIENT AND THE HELIX STILL WOULD NOT EXTEND. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD NVY MEDTRONIC PUERTO RICO, INC. 6947M

Patients

Seq Age Sex Outcome Treatment
1 52 YR Other