FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 2791650 · Received October 15, 2012

Report

Report Number
2182208-2012-03394
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
August 15, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003/S065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE REPORTED ECG (ELECTROCARDIOGRAM) PORT ISSUE; ECG PORT IS BROKEN LOOSE FROM A PRINTED CIRCUIT BOARD ASSEMBLY. SYSTEM FAN IS NOISY. NO ISSUE FOUND WITH THE FLOPPY DRIVE DOOR, WHEN LATCHED IT STAYS CLOSED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ELECTROCARDIOGRAM (ECG) PORT ON THE PROGRAMMER WAS NOT WORKING AND THAT THERE WAS A LOT OF NOISE ON THE ECG. SWAPPING THE CABLES DID NOT RESOLVE BUT TOUCHING OR MOVING THE ECG PORT CLEARED THE NOISE. ALSO, THE SIDE DOOR WHERE THE FLOPPY IS WILL NOT STAY CLOSED. THE PROGRAMMER WAS RETURNED FOR SERVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ELECTROCARDIOGRAM (ECG) PORT ON THE PROGRAMMER WAS NOT WORKING AND THAT THERE WAS A LOT OF NOISE ON THE ECG. SWAPPING THE CABLES DID NOT RESOLVE BUT TOUCHING OR MOVING THE ECG PORT CLEARED THE NOISE. ALSO, THE SIDE DOOR WHERE THE FLOPPY IS WILL NOT STAY CLOSED. THE PROGRAMMER WAS RETURNED FOR SERVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAMMER KRG MEDTRONIC, INC. 2090

Patients

Seq Age Sex Outcome Treatment
1 Other 2067 PROGRAMMER RF HEAD| 2067 PROGRAMMER RF HEAD