FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX

MDR report key: 2791648 · Received October 15, 2012

Report

Report Number
2649622-2012-14853
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
August 20, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S2
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, AFTER MULTIPLE EXTENSIONS AND RETRACTIONS OF THE LEAD HELIX, THE HELIX FAILED TO EXTEND ANYMORE. THEREFORE, THE LEAD WAS NOT USED AND WAS REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5568

Patients

Seq Age Sex Outcome Treatment
1 81 YR Other