Report Number
2649622-2012-15452
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE MAY HAVE REACHED ITS RECOMMENDED REPLACEMENT TIME [RRT] PREMATURELY. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR [RV] LEAD HAD HIGH THRESHOLD MEASUREMENTS FOR THE PAST SIX TO EIGHT MONTHS AND THIS MAY HAVE CONTRIBUTED TO THE PREMATURE DEPLETION OF THE DEVICE. THE DEVICE WAS EXPLANTED AND REPLACED; THE RV LEAD REMAINS IN USE AS IT TESTED "FINE UPON REPLACEMENT." NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076

Patients

Seq Age Sex Outcome Treatment
1 88 YR Hospitalization| R