FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2791609 · Received October 15, 2012

Report

Report Number
2649622-2012-14543
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM AFTER RECEIVING SEVERAL INAPPROPRIATE SHOCKS DUE TO OVERSENSING AND NOISE ON THE RIGHT VENTRICULAR (RV) LEAD. THE PATIENT STATED THAT THE DEVICE HAD BEEN BEEPING FOR A WEEK. IT WAS ALSO REPORTED THAT THE LEAD INTEGRITY ALERT (LIA) HAD TRIGGERED AND THE LEAD HAD HIGH IMPEDANCE AND WAS UNABLE TO CAPTURE. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. THE PATIENT WAS TRANSFERRED TO A TERTIARY FACILITY. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM AFTER RECEIVING SEVERAL INAPPROPRIATE SHOCKS DUE TO OVERSENSING AND NOISE ON THE RIGHT VENTRICULAR (RV) LEAD. THE PATIENT STATED THAT THE DEVICE HAD BEEN BEEPING FOR A WEEK. IT WAS ALSO REPORTED THAT THE LEAD INTEGRITY ALERT (LIA) HAD TRIGGERED AND THE LEAD HAD HIGH IMPEDANCE AND WAS UNABLE TO CAPTURE. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. THE PATIENT WAS TRANSFERRED TO A TERTIARY FACILITY. IT WAS FURTHER REPORTED THAT THE LEAD WAS EXPLANTED DUE TO FRACTURE. THE PATIENT HAD A SUBCLAVIAN PERFORATION AND UNDERWENT AN OPEN CHEST REPAIR OF THE VEIN. THE LEAD WAS REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| L| R 4591 COMPETITOR IMPLANTABLE PACING LEAD| 4591 COMPETITOR IMPLANTABLE PACING LEAD| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4472 COMPETITOR IMPLANTABLE PACING LEAD| 4472 COMPETITOR IMPLANTABLE PACING LEAD