Summary report: N
ADAPTA DR
MDR report key: 2791588
·
Received October 15, 2012
Report
- Report Number
- 6000144-2012-05530
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NVZ
- PMA / PMN Number
- P980035/S043
- Removal / Correction Number
- Z-2120-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE SET SCREW HAD A ROUNDED SOCKET.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ATRIAL LEAD DISLODGED DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE. THE LEAD WAS CAPPED AND A NEW LEAD WAS IMPLANTED. THE DEVICE SET SCREW ON THE ATRIAL CHANNEL WAS STRIPPED AND COULD NOT HOLD THE NEW LEAD. THE DEVICE WAS REMOVED AND REPLACED WITH A NEW DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADAPTA DR | IMPLANTABLE PULSE GENERATOR | NVZ | MEDTRONIC MED REL, INC. | ADDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Hospitalization| O| R | 4076 IMPLANTABLE PACING LEAD |