FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2791543 · Received October 15, 2012

Report

Report Number
2649622-2012-15448
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE PROXIMAL CONDUCTOR WAS FRACTURED. THE OUTER TUBING OVERLAY HAD BLOOD/BODY FLUID, WAS MELTED, AND HAD COSMETIC ENVIRONMENTAL STRESS CRACKING. THE OUTER INSULATION HAD COSMETIC DEPRESSION. THE HELIX/LOBE WAS DISTORTED/BENT. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE NOTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD