FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 2791490 · Received October 15, 2012

Report

Report Number
2182208-2012-03522
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 18, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003/S065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS COULD NOT CONFIRM THE REPORT THAT THE PROGRAMMER HAD NO TELEMETRY. BOTH OF THE LATCHES ON THE POWER CORD DOOR ARE BROKEN. MEDIA BAY DOOR LATCH IS MISSING. (B)(4) RF (RADIO FREQUENCY) HEAD UPLINK AMPLITUDE TESTS ARE OUT OF SPECIFICATION; RF HEAD CABLE FOUND OUT OF ELECTRICAL SPECIFICATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER HAD NO TELEMETRY, AND IT WAS REQUESTED THAT THE WAND, ANALYZER AND PEN BE REPLACED, AND THAT THE PROGRAMMER BE UPGRADED. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS INDICATED ON THE RETURN PAPERWORK THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAMMER RF HEAD KRG MEDTRONIC, INC. 2067

Patients

Seq Age Sex Outcome Treatment
1 Other 2290 PACING SYSTEM ANALYZER