FDA Adverse Event
Malfunction
Summary report: N
CARELINK PROGRAM W/TELEMETRY C
MDR report key: 2791482
·
Received October 15, 2012
Report
- Report Number
- 2182208-2012-03520
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- July 16, 2012
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS CONFIRMED THE REPORTED EVENT, THE LOW VOLTAGE DIFFERENTIAL SIGNALING PRINTED CIRCUIT BOARD WAS LOOSE. IT WAS ALSO NOTED THAT THE RIGHT ANTENNA WAS BROKEN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DISPLAY SCREEN OF THE PROGRAMMER WAS WAVY, AND WHEN IT WAS TURNED OFF, THE PROGRAMMER WOULD NOT REBOOT. IT WAS RETURNED FOR SERVICE. IT WAS INDICATED ON THE RETURN PAPERWORK THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK PROGRAM W/TELEMETRY C | PROGRAMMER | DXY | MEDTRONIC, INC. | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |