FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2791482 · Received October 15, 2012

Report

Report Number
2182208-2012-03520
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 16, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS CONFIRMED THE REPORTED EVENT, THE LOW VOLTAGE DIFFERENTIAL SIGNALING PRINTED CIRCUIT BOARD WAS LOOSE. IT WAS ALSO NOTED THAT THE RIGHT ANTENNA WAS BROKEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DISPLAY SCREEN OF THE PROGRAMMER WAS WAVY, AND WHEN IT WAS TURNED OFF, THE PROGRAMMER WOULD NOT REBOOT. IT WAS RETURNED FOR SERVICE. IT WAS INDICATED ON THE RETURN PAPERWORK THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1 Other