FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2791478 · Received October 15, 2012

Report

Report Number
2649622-2012-14529
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 10, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4) : THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THERE WAS BLOOD/BODY FLUID (NOT OBSTRUCTED) ON THE PROXIMAL CONDUCTOR AND A BREACHED CUT ON THE OUTER INSULATION. THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE RIGHT VENTRICULAR LEAD HELIX WOULD NOT EXTEND. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI

Patients

Seq Age Sex Outcome Treatment
1 75 YR Other