Malfunction
Summary report: N
CAPSUREFIX NOVUS MRI
MDR report key: 2791478
·
Received October 15, 2012
Report
- Report Number
- 2649622-2012-14529
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- July 10, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4) : THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THERE WAS BLOOD/BODY FLUID (NOT OBSTRUCTED) ON THE PROXIMAL CONDUCTOR AND A BREACHED CUT ON THE OUTER INSULATION. THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND APPARENT EXPLANT DAMAGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE RIGHT VENTRICULAR LEAD HELIX WOULD NOT EXTEND. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Other |