FDA Adverse Event Injury Summary report: N

MAXIMO II CRT-D

MDR report key: 2791435 · Received October 15, 2012

Report

Report Number
6000144-2012-05347
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
Z-0112-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE LASTED SEVENTY-FOUR PERCENT OF ITS EXPECTED LONGEVITY. BOTH LOADED AND UNLOADED PACING CURRENT DRAIN LEVELS WERE NORMAL FOR THIS DEVICE OVER THE RANGE OF BATTERY VOLTAGE IT OPERATED UNDER DURING ITS SERVICE TIME. THERE IS NO EVIDENCE TO INDICATE A PROBLEM WITH THE BATTERY. WITHOUT KNOWING THE PROGRAMMING HISTORY OF THIS DEVICE THERE IS NO WAY TO DETERMINE WHY IT DID NOT MEET ITS EXPECTED LONGEVITY CALCULATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD EARLY BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAXIMO II CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D284TRK

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| R 4196 IMPLANTABLE PACING LEAD| 0145 COMPETITOR IMPLANTABLE TACHY LEAD| 4087 COMPETITOR IMPLANTABLE PACING LEAD