MAXIMO II CRT-D
Report
- Report Number
- 6000144-2012-05347
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S084
- Removal / Correction Number
- Z-0112-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE LASTED SEVENTY-FOUR PERCENT OF ITS EXPECTED LONGEVITY. BOTH LOADED AND UNLOADED PACING CURRENT DRAIN LEVELS WERE NORMAL FOR THIS DEVICE OVER THE RANGE OF BATTERY VOLTAGE IT OPERATED UNDER DURING ITS SERVICE TIME. THERE IS NO EVIDENCE TO INDICATE A PROBLEM WITH THE BATTERY. WITHOUT KNOWING THE PROGRAMMING HISTORY OF THIS DEVICE THERE IS NO WAY TO DETERMINE WHY IT DID NOT MEET ITS EXPECTED LONGEVITY CALCULATION.
IT WAS REPORTED THAT THE DEVICE HAD EARLY BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO II CRT-D | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | D284TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Hospitalization| R | 4196 IMPLANTABLE PACING LEAD| 0145 COMPETITOR IMPLANTABLE TACHY LEAD| 4087 COMPETITOR IMPLANTABLE PACING LEAD |