FDA Adverse Event Malfunction Summary report: N

SPRINT FIDELIS

MDR report key: 2791425 · Received October 15, 2012

Report

Report Number
2649622-2012-14524
Event Type
Malfunction
Date Received
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0070-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): PERFORMANCE DATA WERE COLLECTED FROM THE DEVICE. THE WEEKLY PACING LEAD MEASUREMENT LOG DATA SHOWS A GRADUAL INCREASE FOR THE MINIMUM AND MAXIMUM RIGHT VENTRICULAR PACING IMPEDANCE MEASUREMENTS FROM 400 TO 584 OHMS BETWEEN (B)(4) 2011 AND (B)(4) 2012.

Description of Event or Problem · 1

IT WAS REPORTED THE LEAD IMPEDANCE MEASUREMENTS HAVE SLOWLY INCREASED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6949

Patients

Seq Age Sex Outcome Treatment
1 60 YR Other 7304 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD