FDA Adverse Event
Malfunction
Summary report: N
SPRINT FIDELIS
MDR report key: 2791425
·
Received October 15, 2012
Report
- Report Number
- 2649622-2012-14524
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S029
- Removal / Correction Number
- Z-0070-2008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): PERFORMANCE DATA WERE COLLECTED FROM THE DEVICE. THE WEEKLY PACING LEAD MEASUREMENT LOG DATA SHOWS A GRADUAL INCREASE FOR THE MINIMUM AND MAXIMUM RIGHT VENTRICULAR PACING IMPEDANCE MEASUREMENTS FROM 400 TO 584 OHMS BETWEEN (B)(4) 2011 AND (B)(4) 2012.
Description of Event or Problem · 1
IT WAS REPORTED THE LEAD IMPEDANCE MEASUREMENTS HAVE SLOWLY INCREASED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT FIDELIS | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6949 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Other | 7304 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD |