FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2791398 · Received October 15, 2012

Report

Report Number
1416980-2012-02431
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
January 11, 2012
Report Date
January 11, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE CUSTOMER REPORTED PROBLEM OF "ALARM F-49" WAS CONFIRMED DURING DEVICE EVALUATION. SERVICE DETERMINED THAT THE CAUSE WAS DUE OUTDATED BATTERIES, WHICH WERE REPLACED TO RESOLVE THE ISSUE. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A FAILURE CODE OF 49. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THE CUSTOMER DID NOT REPORT IF THERE WAS ANY PATIENT INVOLVEMENT; THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1