FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 2791369 · Received October 15, 2012

Report

Report Number
2182208-2012-03512
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 10, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
NVZ
PMA / PMN Number
P890003/S145
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS FOUND THAT THE POWER SUPPLY WAS OUT OF SPECIFICATION, DUE TO THIS CONDITION THE PROGRAMMER WAS UNABLE TO POWER UP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER WOULD NOT POWER ON, DESPITE CHANGING OUT THE POWER CORD AND TRYING A DIFFERENT POWER OUTLET. IT WAS FURTHER REPORTED THAT THE FANS ALSO DID NOT TURN ON. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS ALSO INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAMMER NVZ MEDTRONIC, INC. 2090X

Patients

Seq Age Sex Outcome Treatment
1 Other 2067 PROGRAMMER RF HEAD