REVO MRI SURESCAN
Report
- Report Number
- 6000144-2012-05633
- Event Type
- Injury
- Date Received
- October 15, 2012
- Date of Event
- August 9, 2012
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWP
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE INNER INSULATION WAS KINKED/BUCKLED, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED. (B)(4): THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THE PATIENT DEVELOPED AN INFECTION. THE DEVICE AND LEADS WERE REMOVED AND THE SYSTEM WILL BE REPLACED AT A FUTURE DATE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REVO MRI SURESCAN | IMPLANTABLE PULSE GENERATOR | LWP | MEDTRONIC MED REL, INC. | RVDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Hospitalization| R |