FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2791326 · Received October 15, 2012

Report

Report Number
2649622-2012-15429
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
August 22, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVY
PMA / PMN Number
P920015/S055
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY : (B)(4) : THE FULL LEAD WAS RETURNED AND ANALYZED; ANALYSIS REVEALED THE HELIX WAS BLOCKED BY THE GUIDE TOOTH. THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THE LEAD WAS DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE LEAD WAS ATTEMPTED BUT NOT USED BECAUSE THE HELIX WOULD NOT EXTEND. ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD NVY MEDTRONIC PUERTO RICO, INC. 6947M

Patients

Seq Age Sex Outcome Treatment
1 Other