FDA Adverse Event Injury Summary report: N

ENPULSE

MDR report key: 2791320 · Received October 15, 2012

Report

Report Number
6000094-2012-02156
Event Type
Injury
Date Received
October 15, 2012
Report Date
November 30, 2012
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035/S032
Removal / Correction Number
Z-2120-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) - THE DEVICE WAS RETURNED AND ANALYZED. THE BATTERY DEPLETION WAS FOUND TO BE NORMAL AND MET EXPECTED LONGEVITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE PHYSICIAN THOUGHT THE BATTERY SEEMED TO HAVE DROPPED PREMATURELY FROM THE LAST FOLLOW UP. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENPULSE PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND E2DR01

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R