FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO

MDR report key: 2791285 · Received October 15, 2012

Report

Report Number
2649622-2012-15425
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
August 27, 2012
Report Date
December 13, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015/S24
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THERE WERE NO ANOMALIES FOUND. (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE LEAD WAS OPENED BUT NOT IMPLANTED. THE PHYSICIAN CHECKED THE HELIX OF THE LEAD BEFORE IMPLANT AND ROTATED IT 30 TIMES AS INDICATED IN THE MANUAL. THE HELIX DID NOT EXTEND AND THE DOCTOR DID NOT FEEL COMFORTABLE USING THIS LEAD. ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6947

Patients

Seq Age Sex Outcome Treatment
1 70 YR Other