FDA Adverse Event
Malfunction
Summary report: N
CARELINK PROGRAM W/TELEMETRY C
MDR report key: 2791273
·
Received October 15, 2012
Report
- Report Number
- 2182208-2012-03504
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- July 12, 2012
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE PROGRAMMER WAS RETURNED AND ANALYSIS COULD NOT CONFIRM THE CUSTOMER COMMENT THAT THE DEVICE WOULD OCCASIONALLY NOT POWER ON. THE DEVICE WAS TURNED OFF AND BACK ON AND FUNCTIONALLY TESTED AND NO FAULT WAS FOUND. A STAPLE WAS FOUND IN THE UNIVERSAL SERIAL BUS (USB) PORT. THERE WAS A NOISY SYSTEM FAN WHICH WAS REPLACED AND A BROKEN RIGHT KEYBOARD HINGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PROGRAMMER WOULD OCCASIONALLY NOT POWER UP. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT INVOLVEMENT WAS INDICATED FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK PROGRAM W/TELEMETRY C | PROGRAMMER | DXY | MEDTRONIC, INC. | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | 2067 PROGRAMMER RF HEAD |