FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2791273 · Received October 15, 2012

Report

Report Number
2182208-2012-03504
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 12, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE PROGRAMMER WAS RETURNED AND ANALYSIS COULD NOT CONFIRM THE CUSTOMER COMMENT THAT THE DEVICE WOULD OCCASIONALLY NOT POWER ON. THE DEVICE WAS TURNED OFF AND BACK ON AND FUNCTIONALLY TESTED AND NO FAULT WAS FOUND. A STAPLE WAS FOUND IN THE UNIVERSAL SERIAL BUS (USB) PORT. THERE WAS A NOISY SYSTEM FAN WHICH WAS REPLACED AND A BROKEN RIGHT KEYBOARD HINGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER WOULD OCCASIONALLY NOT POWER UP. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT INVOLVEMENT WAS INDICATED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1 Other 2067 PROGRAMMER RF HEAD