FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2791226 · Received October 15, 2012

Report

Report Number
2649622-2012-15419
Event Type
Injury
Date Received
October 15, 2012
Date of Event
August 30, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): NO ANOMALIES FOUND, HOWEVER, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THE NUMBER OF TURNS TO EXTEND/RETRACT THE HELIX EXCEED SPECIFICATION AND THE LEAD APPEARED DAMAGED AT IMPLANT; FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED A POSSIBLE PERICARDIAL EFFUSION. THE ATRIAL LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| L| R 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR