CONCERTO II CRT-D
Report
- Report Number
- 6000094-2012-02282
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- NIK
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY: (B)(4) - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. HIGH RESISTANCE/IMPEDANCE WAS NOTED AS TWO PATIENT ALERTS FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE OCCURRED ON (B)(4) 2012 15:00:09 AND (B)(4) 2012 09:00:12. THE WEEKLY PACE LEAD IMPEDANCE TREND DATA SHOW VARIOUS SPIKE INCREASES FOR MAX VENTRICULAR PACE BI IMPEDANCE FROM 741 TO 4047 OHMS RANGE BETWEEN (B)(4) 2012. OVERSENSING WAS NOTED AS 14 VENTRICULAR NON SUSTAINED TACHYCARDIA EPISODES <=210 MS OCCURRED BETWEEN (B)(4) 2012 22:27:14 AND (B)(4) 2012 06:38:07. LEAD INTEGRITY ALERT WAS TRIGGERED AS TWO PATIENT ALERTS FOR LEAD FAILURE PREDICTOR OCCURRED ON (B)(4) 2012 15:40:06 AND (B)(4) 2012 16:49:29. (B)(4). THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA (B)(4) - COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. (B)(4) - IMPEDANCE - HIGH RESISTANCE/IMPEDANCE (B)(4) 2 - PATIENT ALERTS FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(4) 2012 15:00:09 AND (B)(4) 2012 09:00:12. WEEKLY PACE LEAD IMPEDANCE TREND DATA SHOW VARIOUS SPIKE INCREASES FOR MAX VENTRICULAR PACE BI IMPEDANCE = 741 TO 4047 OHMS RANGE BETWEEN (B)(4) 2012. (B)(4) SENSING - OVERSENSING (B)(4) 14 - VENTRICULAR NST<=210 MS BETWEEN (B)(4) 2012 22:27:14 AND (B)(4) 2012 06:38:07. (B)(4) STD REVIEW - LEAD INTEGRITY ALERT TRIGGERED (B)(4) 2 - PATIENT ALERTS FOR LEAD FAILURE PREDICTOR ON (B)(4)2012 15:40:06 AND (B)(4) 2012 16:49:29.
IT WAS REPORTED THAT A LEAD INTEGRITY ALERT WAS TRIGGERED. DURING THE REVISION PROCEDURE, A LOOSE CONNECTION WAS OBSERVED AND CORRECTED. APPROXIMATELY THREE MONTHS LATER, ANOTHER LEAD INTEGRITY ALERT WAS TRIGGERED AND THE RIGHT VENTRICULAR LEAD HAD HIGH IMPEDANCE. THE DEVICE AND LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS WERE NOTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO II CRT-D | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC S.A. | D294TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| O| R | 5594 IMPLANTABLE PACING LEAD| 4193 IMPLANTABLE PACING LEAD |