CAPSUREFIX NOVUS

MDR report key: 2791183 · Received October 15, 2012

Report

Report Number
2649622-2012-15413
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR WAS FRACTURED. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THE OUTER INSULATION WAS BREACHED CUT, THE HELIX/LOBE WAS DISTORTED/BENT AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT FELT DIZZY AND PRESENTED TO THE HOSPITAL. ELECTROCARDIOGRAPHY REVEALED NO PACING AND THE PHYSICIAN FELT THIS WAS DUE TO A FRACTURE OF THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE NOTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076

Patients

Seq Age Sex Outcome Treatment
1 93 YR Hospitalization| R ADSR01 IMPLANTABLE PULSE GENERATOR