CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2012-14503
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- August 6, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. BLOOD/BODY FLUID WAS FOUND ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED) AND THE OUTER INSULATION WAS BREACHED CUT. BLOOD WAS FOUND IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.
IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE RIGHT VENTRICULAR [RV] LEAD WAS REPOSITIONED MULTIPLE TIMES BUT HAD POOR SENSING AND HIGH THRESHOLD MEASUREMENTS. THE PHYSICIAN REMOVED THE LEAD STATING THAT THERE WERE "POSSIBLE PROBLEMS." A DIFFERENT RV LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Other |