FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2791162 · Received October 15, 2012

Report

Report Number
2649622-2012-14503
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
August 6, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. BLOOD/BODY FLUID WAS FOUND ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED) AND THE OUTER INSULATION WAS BREACHED CUT. BLOOD WAS FOUND IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE RIGHT VENTRICULAR [RV] LEAD WAS REPOSITIONED MULTIPLE TIMES BUT HAD POOR SENSING AND HIGH THRESHOLD MEASUREMENTS. THE PHYSICIAN REMOVED THE LEAD STATING THAT THERE WERE "POSSIBLE PROBLEMS." A DIFFERENT RV LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076

Patients

Seq Age Sex Outcome Treatment
1 63 YR Other