FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2791158 · Received October 15, 2012

Report

Report Number
2649622-2012-15101
Event Type
Malfunction
Date Received
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE REMOTE TRANSMISSION SHOWED AN EPISODE WITH A T-WAVE OVERSENSING (TWOS). IT WAS INTERMITTENT AT FIRST AND COLLECTED BOTH A HIGH RATE NON-SUSTAINED AND ALSO A VENTRICULAR FIBRILLATION (VF) EPISODE THAT WERE WITHHELD FOR TWOS. THE LEAD WILL CONTINUE TO BE MONITORED, AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other D314TRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4196 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD