FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2791144 · Received October 15, 2012

Report

Report Number
2182208-2012-03492
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 6, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4). ANALYSIS CONFIRMED THE REPORTED EVENT. STYLUS IS INTERMITTENT. CONNECTOR IS DAMAGED/BENT. HINGE PLATE IS CRACKED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PEN (STYLUS) WAS NOT WORKING PROPERLY. IT WAS FURTHER NOTED THE "PEN INPUT CONNECTOR FAULTY". THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1 Other 2290 PACING SYSTEM ANALYZER