FDA Adverse Event
Malfunction
Summary report: N
CARELINK PROGRAM W/TELEMETRY C
MDR report key: 2791144
·
Received October 15, 2012
Report
- Report Number
- 2182208-2012-03492
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- July 6, 2012
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4). ANALYSIS CONFIRMED THE REPORTED EVENT. STYLUS IS INTERMITTENT. CONNECTOR IS DAMAGED/BENT. HINGE PLATE IS CRACKED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PEN (STYLUS) WAS NOT WORKING PROPERLY. IT WAS FURTHER NOTED THE "PEN INPUT CONNECTOR FAULTY". THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK PROGRAM W/TELEMETRY C | PROGRAMMER | DXY | MEDTRONIC, INC. | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | 2290 PACING SYSTEM ANALYZER |