FDA Adverse Event Injury Summary report: N

SCREW-IN

MDR report key: 2791140 · Received October 15, 2012

Report

Report Number
2649622-2012-15100
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
K902002
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION AND THE DEVICE MET EXPECTED LONGEVITY. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. TIME OF RECOMMENDED REPLACEMENT TIME (RRT) IN SAVE TO DISK OCCURRED ON (B)(4) 2012 WITH DEVICE RRT <=2.6251 VOLT. WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MIN BAT=2.632 TO 2.618 VOLTS BETWEEN (B)(4) 2012. A PATIENT ALERT FOR LOW BATTERY VOLTAGE OCCURRED ON (B)(4) 2012 02:15:00. (B)(4) - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD HIGH THRESHOLD. IT WAS ALSO REPORTED THAT THE IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) REACHED EARLY ELECTIVE REPLACEMENT INDICATOR (ERI). THE LV LEAD AND ICD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SCREW-IN IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5071

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization| R 1688TC COMPETITOR IMPLANTABLE PACING LEAD| 6944 IMPLANTABLE TACHY LEAD