MAXIMO II VR
Report
- Report Number
- 6000094-2012-02194
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S114
- Removal / Correction Number
- Z-0114-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. SENSING - OVERSENSING: THERE WERE THIRTEEN VENTRICULAR NST<=210 MS BETWEEN (B)(6) 2012 02:58:48 AND (B)(6) 2012 12:14:32. SENSING - INTERFERENCE/NOISE: VENTRICULAR SHORT INTERVAL COUNT V-SIC=287.5 COUNTS AVG/DAY, IN 58.11 DAYS, BETWEEN (B)(6) 2012 10:26:30 AND (B)(6) 2012 13:00:55. EVALUATION SUMMARY (B)(4): THE DEVICE WAS RETURNED TO THE MANUFACTURER AND ANALYZED AND NO ANOMALIES WERE FOUND.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD LOW R-WAVES. THEREFORE THE PACE/SENSE PORTION OF THE RV LEAD WAS CAPPED AND REPLACED WITH A NEW PACE/SENSE LEAD. THE HIGH VOLTAGE PORTION OR THE RV LEAD REMAINS IN USE. IT WAS ALSO REPORTED THAT IT WAS DISCOVERED THAT THE NEW PACE/SENSE LEAD WASN'T PROPERLY PLACED IN THE RV PORT OF THE DEVICE AND THAT IT WAS IMPOSSIBLE TO PULL THE NEW PACE/SENSE LEAD OUT OF THE RV PORT. THEREFORE THE DEVICE WAS REMOVED AND REPLACED WITH A NEW DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO II VR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC S.A. | D284VRC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Hospitalization| R | SR60 COMPETITOR IMPLANTABLE PACING LEAD |