FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 2791123 · Received October 15, 2012

Report

Report Number
2182208-2012-03487
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
July 5, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
KRG
PMA / PMN Number
P890003/S065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PROGRAMMER HAD INTERMITTENT TELEMETRY WITH THE RF (RADIO FREQUENCY) HEAD; THE RF HEAD CABLE FOUND OUT OF ELECTRICAL SPECIFICATION.

Description of Event or Problem · 1

THE RF (RADIO FREQUENCY) HEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAMMER RF HEAD KRG MEDTRONIC, INC. 2067

Patients

Seq Age Sex Outcome Treatment
1 Other (B)(4) PROGRAMMER