FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS MRI
MDR report key: 2791111
·
Received October 15, 2012
Report
- Report Number
- 2649622-2012-14203
- Event Type
- Injury
- Date Received
- October 15, 2012
- Date of Event
- August 4, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT TWO DAYS AFTER IMPLANT, THE PATIENT CAME TO THE EMERGENCY ROOM REPORTING TO HAVE A LOW HEART RATE. NO LOSS OF CAPTURE OR BRADYCARDIA WERE FOUND. THE SYSTEM TESTED NORMAL. THE PHYSICIAN OPTED TO REPLACE THE RIGHT ATRIAL LEAD FOR SUSPECTED LOSS OF CAPTURE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR |