FDA Adverse Event
Injury
Summary report: N
SENSIA SR
MDR report key: 2791110
·
Received October 15, 2012
Report
- Report Number
- 6000094-2012-02193
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- NVZ
- PMA / PMN Number
- P980035/S043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR: (B)(4) - THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) HAD TO BE REPLACED BECAUSE THE BATTERY RAN OUT BEFORE THE EXPECTED TIME. THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SENSIA SR | IMPLANTABLE PULSE GENERATOR | NVZ | MEDTRONIC S.A. | SESR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |