SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2012-14476
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DEFIBRILLATION CONDUCTOR WAS DISTORTED, THE OUTER INSULATION HAD COSMETIC DEPRESSION, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, TISSUE ON THE HELIX, AND VISUAL ANALYSIS REVEALED APPARENT EXPLANT DAMAGE. ANALYST VISUAL COMMENT, THE HELIX IS BENT, BUT OPERATES WITHIN SPECIFICATION. IT CANNOT BE DETERMINED AT THIS TIME IF THE HELIX GOT BENT DURING IMPLANT OR EXPLANT.
AS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD R WAVES HAVE DIMINISHED. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |