FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 2790929
·
Received October 15, 2012
Report
- Report Number
- 2649622-2012-14474
- Event Type
- Injury
- Date Received
- October 15, 2012
- Date of Event
- August 9, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD HIGH IMPEDANCE AND THRESHOLD, OVERSENSING, AND THAT THE LEAD INTEGRITY ALERT HAD TRIPPED. THE PATIENT WAS TAKEN BACK TO SURGERY AND IT WAS NOTED ON FLUORO THAT THE PIN IN THE RV LEAD WAS NOT EXTENDED PAST THE SETSCREW. THE PHYSICIAN REINSERTED THE RV LEAD PIN AND FOUND THE PIN TO BE SECURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4193 IMPLANTABLE PACING LEAD |