FDA Adverse Event Injury Summary report: N

ADAPTA DR

MDR report key: 2790887 · Received October 15, 2012

Report

Report Number
6000144-2012-05717
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NVZ
PMA / PMN Number
P980035/S043
Removal / Correction Number
Z-2120-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4): THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE INNER INSULATION WAS KINKED/BUCKLED, THE OUTER INSULATION WAS BREACH CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED. (B)(4): THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS DISTORTED, THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACH CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), THERE WAS TISSUE ON THE HELIX AND THERE WAS APPARENT EXPLANT DAMAGE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT DEVELOPED "PACEMAKER ENDOCARDITIS". THE DEVICE AND LEADS WERE REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADAPTA DR IMPLANTABLE PULSE GENERATOR NVZ MEDTRONIC MED REL, INC. ADDR01

Patients

Seq Age Sex Outcome Treatment
1 89 YR Hospitalization| L| R