SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2012-14458
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. SENSING - OVERSENSING. 1 - VENTRICULAR NST=190 MS BETWEEN (B)(4) 2012 08:27:44. (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA (B)(4) COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA.(B)(4) SENSING - OVERSENSING (B)(4) 1 - VENTRICULAR NST=190 MS BETWEEN (B)(4) 2012 08:27:44.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS OVERSENSING NOISE. THE PATIENT WILL BE MONITORED AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Other | 4196 IMPLANTABLE PACING LEAD| 4076 IMPLANTABLE PACING LEAD| D274TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB |