FDA Adverse Event Injury Summary report: N

CAPSURE SENSE

MDR report key: 2790765 · Received October 15, 2012

Report

Report Number
2649622-2012-14454
Event Type
Injury
Date Received
October 15, 2012
Date of Event
July 30, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S034
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR [RV] LEAD HAD NO CAPTURE AT MAXIMUM OUTPUT AND WAS INTERMITTENTLY UNDERSENSING DUE TO LEAD DISLODGEMENT. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SENSE IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4074

Patients

Seq Age Sex Outcome Treatment
1 92 YR Hospitalization| R SESR01 IMPLANTABLE PULSE GENERATOR