SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2012-14765
- Event Type
- Injury
- Date Received
- October 15, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
PRODUCT EVALUATION SUMMARY: THE FULL LENGTH OF THE LEAD IN SEGMENTS WAS RETURNED, ANALYZED AND THE HELIX WAS FRACTURED FROM BEING PULLED/STRETCHED/OVERSTRESSED. IT WAS NOTED THAT THE DISTAL CONDUCTOR HAD BLOOD (NOT OBSTRUCTED) AND THE EXPOSED SUPERIOR VENA CAVA (SVC) AND RIGHT VENTRICULAR (RV) DEFIBRILLATION CONDUCTOR COILS WERE BOTH KINKED/BUCKLED. THE DISTAL END LOW VOLTAGE ELECTRODE WAS COVERED WITH BLOOD AND BODY TISSUE/FIBROTIC GROWTH AND WAS DAMAGED FROM BEING PULLED/STRETCHED/OVERSTRESSED. THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING AND THE OUTER TUBING OVER LAY WAS BREACHED FROM BEING MELTED. THE LEAD HAD APPARENT EXPLANT DAMAGE. THE ANALYST COMMENTED THAT PER SCANNING ELECTRON MICROSCOPY, THE HELIX WIRE FRACTURED BY AN OVERLOAD CONDITION WITH A TORSIONAL COMPONENT WHICH IS INDICATED BY THE STEP IN THE FRACTURE FACE. THE HELIX, AT THE SHANK, IS SUPPOSED TO HAVE APPROXIMATELY 2.5 WIRE TURNS AND THIS LEAD HAS NONE, SO THE WELD WAS PROBABLY ALSO CLOSE TO FRACTURING. THE HELIX WIRE IS ALSO STRETCHED OUT INDICATING A LARGE TENSILE LOAD WAS APPLIED (WHICH TRANSLATES INTO A TORSION FOR A COIL). IT WAS TRIED TO ESTIMATE HOW MUCH OF THE HELIX IS MISSING AND APPROXIMATELY 1 AND 1/2 TURNS OF THE WIRE ARE MISSING.
IT WAS REPORTED THAT THERE WAS T-WAVE OVERSENSING AND AN INCREASE IN RIGHT VENTRICULAR PACING IMPEDANCE FROM 700 OHMS TO 2000 OHMS. IT WAS ALSO REPORTED THAT THE UNSTABLE IMPEDANCES OCCURRED OVER SEVERAL WEEKS. IT WAS FURTHER REPORTED THAT THERE WAS AN INCREASE IN THRESHOLDS. THE SENSITIVITY WAS CHANGED. THE PATIENT WILL BE SCHEDULED FOR FURTHER EVALUATION. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THERE WAS T-WAVE OVERSENSING AND AN INCREASE IN RIGHT VENTRICULAR PACING IMPEDANCE FROM 700 OHMS TO 2000 OHMS. IT WAS ALSO REPORTED THAT THE UNSTABLE IMPEDANCES OCCURRED OVER SEVERAL WEEKS. IT WAS ALSO REPORTED THAT THERE WAS AN INCREASE IN THRESHOLDS. THE SENSITIVITY WAS CHANGED. IT WAS FURTHER REPORTED THAT DURING THE EXTRACTION OF THE LEAD AN ATTEMPT TO RETRACT THE HELICAL SCREW WAS NOT SUCCESSFUL. THEREFORE THE VERY TIP OF THE HELICAL SCREW WAS LEFT IN THE MYOCARDIUM. IT WAS NOTED THAT THIS LIKELY REPRESENTS THE EXPLANATION FOR THE HIGH IMPEDANCE AND CAPTURE THRESHOLDS ASSOCIATED WITH THE LEAD. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Hospitalization| R | D284DRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| D284DRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD |