FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 2790727
·
Received October 15, 2012
Report
- Report Number
- 2649622-2012-14447
- Event Type
- Injury
- Date Received
- October 15, 2012
- Date of Event
- August 8, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVY
- PMA / PMN Number
- P920015/S055
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE AND DEFIBRILLATION TESTING THE PHYSICIAN WAS UNABLE TO SUCCESSFULLY TERMINATE VENTRICULAR FIBRILLATION (VF) WITH ANYTHING LESS THAN 35 JOULES. IT WAS ALSO REPORTED THAT REVERSING THE POLARITY AND TURNING OFF THE SUPERIOR VENA CAVA (SVC) COIL WAS UNSUCCESSFUL. THE LEAD WAS REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | NVY | MEDTRONIC PUERTO RICO, INC. | 6947M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |