FDA Adverse Event Malfunction Summary report: N

EXTERNAL PULSE GENERATOR

MDR report key: 2790620 · Received October 15, 2012

Report

Report Number
2183613-2012-01673
Event Type
Malfunction
Date Received
October 15, 2012
Date of Event
June 13, 2012
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): BATTERY CONTACTS ARE COMPRESSED. ONE PRINTED CIRCUIT BOARD FLEX IS CREASED. UPPER AND LOWER CASES, BOTH BAIL COVERS, RING COVER, BATTERY DRAWER, AND LCD (LIQUID CRYSTAL DISPLAY) ARE BROKEN. LEAD FLEX COVER IS CORRODED. THE RING IS BENT. KEYBOARD WINDOW IS SCRATCHED.

Description of Event or Problem · 1

IT WAS REPORTED THE EPG (EXTERNAL PULSE GENERATOR) TOOK A FALL FROM TABLE HEIGHT. THE EPG WAS RETURNED TO THE MANUFACTURER FOR REPAIR AND CALIBRATION, ANALYZED AND TESTED OUT OF SPECIFICATION. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388

Patients

Seq Age Sex Outcome Treatment
1 Other