FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX

MDR report key: 2790578 · Received October 15, 2012

Report

Report Number
2649622-2012-15040
Event Type
Malfunction
Date Received
October 15, 2012
Report Date
March 15, 2019
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED. ANALYSIS INDICATED THAT THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO ENVIRONMENTAL STRESS CRACKING WHILE IN VIVO. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD SHOWED NOISE OVERSENSING ON AN ATRIAL HIGH RATE EPISODE. THE NOISE WAS REPRODUCIBLE DURING ISOMETRIC TESTING IN THE OFFICE. THE LEAD REMAINS IN USE AND CONTINUES TO BE MONITORED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE DTB MEDTRONIC PUERTO RICO, INC. 4068

Patients

Seq Age Sex Outcome Treatment
1 44 YR Other 5076 IMPLANTABLE PACING LEAD| ADDR01 IMPLANTABLE PULSE GENERATOR