FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX
MDR report key: 2790578
·
Received October 15, 2012
Report
- Report Number
- 2649622-2012-15040
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Report Date
- March 15, 2019
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED. ANALYSIS INDICATED THAT THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO ENVIRONMENTAL STRESS CRACKING WHILE IN VIVO. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ATRIAL LEAD SHOWED NOISE OVERSENSING ON AN ATRIAL HIGH RATE EPISODE. THE NOISE WAS REPRODUCIBLE DURING ISOMETRIC TESTING IN THE OFFICE. THE LEAD REMAINS IN USE AND CONTINUES TO BE MONITORED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | DTB | MEDTRONIC PUERTO RICO, INC. | 4068 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Other | 5076 IMPLANTABLE PACING LEAD| ADDR01 IMPLANTABLE PULSE GENERATOR |