CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2012-14149
- Event Type
- Malfunction
- Date Received
- October 15, 2012
- Date of Event
- August 1, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE U.S. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. HOWEVER, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. VISUAL ANALYSIS REVEALED THE HELIX FUNCTION WAS WITHIN SPECIFICATION.
IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE HELIX COULD NOT BE EXTENDED. THE PHYSICIAN ALSO ATTEMPTED TO EXTEND THE HELIX OUTSIDE THE BODY WITH NO SUCCESS. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |