FDA Adverse Event Injury Summary report: N

KAPPA 900 DR

MDR report key: 2790551 · Received October 15, 2012

Report

Report Number
2647346-2012-01714
Event Type
Injury
Date Received
October 15, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/014
Removal / Correction Number
Z-2120-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS ATTEMPT TO INTERROGATE THE DEVICE BUT TELEMETRY WAS NEVER ESTABLISHED. IT WAS ASSUMED THAT THE BATTERY WAS AT END-OF-LIFE. THE DEVICE WAS REMOVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KAPPA 900 DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. KDR901

Patients

Seq Age Sex Outcome Treatment
1 79 YR Hospitalization| R 4194 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| D314TRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD